10A NCAC 41B .0320. INTOXILYZER: MODEL 5000  


Latest version.
  • The operational procedures to be followed in using the Intoxilyzer, Model 5000 are:

    (1)           Insure instrument displays time and date;

    (2)           Insure observation period requirements have been met;

    (3)           Press "START TEST"; when "INSERT CARD" appears, insert test record;

    (4)           Enter information as prompted;

    (5)           Verify instrument accuracy;

    (6)           When "PLEASE BLOW" appears, collect breath sample;

    (7)           When "PLEASE BLOW" appears, collect breath sample; and

    (8)           When test record ejects, remove.

    If the alcohol concentrations differ by more than 0.02, a third breath sample shall be collected when "PLEASE BLOW" appears.  Subsequent tests shall be administered as soon as feasible by repeating steps (1) through (8), as applicable.

     

History Note:        Authority G.S. 20‑139.1(b);

Eff. January 1, 1985;

Amended Eff. November 1, 2007; April 1, 2001; April 1, 1993; April 1, 1992; January 1, 1990; March 1, 1989.